Universal PCXR8 Sample Pump Single Kit

Product code : 1684295133

The Universal PCXR8 personal or area air sampling pump features a 5 to 5000 ml/min flow range, programmability, and up to 12-hour run time or extended intermittent sampling for particulates and gases/vapors. The pump is ideal for use with sorbent tubes, filter cassettes, size-selective samplers, sample bags, and impingers. The Universal PCXR8 air sampling pump is UL Listed for intrinsic safety.

Quantity:
More information:
  • Nhà Sản Xuât
  • SKC Inc

    Specifications:

    Flow Range

    1000 to 5000 ml/min (UL Listed model)
    (5 to 500 ml/min requires the 224-26 Series adjustable tube holder)
    Requires 1/4-inch ID tubing

    Weight
    34 oz (964 gm)
    Dimensions
    5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.8 cm)
    Compensation Range
    1000 to 2500 ml/min at 40 inches water back pressure
    3000 ml/min at 35 inches water back pressure
    4000 ml/min at 20 inches water back pressure
    5000 ml/min at 10 inches water back pressure
    Typical Back Pressure
    See Table 1 below.
    Flow Control
    Holds constant flow to ± 5% of set-point
    Run Time
    NiCad Battery: 8 hrs minimum at 4000 ml/min and 20 inches water back pressure; dependent on media used. See Table 3.
    NiMH Battery: 12 hrs minimum at 4000 ml/min and 20 inches water back pressure; dependent on media used. See Table 2.
    Battery Eliminator: Pump will provide extended runs.
    Tubing Requires 1/4-inch ID tubing
    Timer Display Range
    1 to 9999 minutes (6.8 days). Pump will shut off at 9999 minutes. To reset, restart the pump.
    Flow Indicator
    Built-in rotameter with 250-ml division; scale marked at 1, 2, 3, 4, and 5 L/min
    Charging Time
    (varies with battery capacity and level of discharge)
    6 to 8.5 hrs with PowerFlex charger
    Power Supply
    6-V plug-in NiMH battery pack, rechargeable, 3.5-Ah capacity or
    6-V plug-in NiCad battery pack, rechargeable, 2.0-Ah capacity
    battery eliminator is available (see Ordering Information); use voids the UL Listing for intrinsic safety.
    Intrinsic Safety

    UL Listed for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and Class III, Temperature Code T3C
    ATEX-approved model is available. Contact SKC.

    Operating Temperature
    32 to 104 F (0 to 40 C)
    Storage Temperature
    -4 to 113 F (-20 to 45 C)
    Charging Temperature
    50 to 113 F (10 to 45 C)
    Operating Humidity
    0 to 95% non-condensing
    Flow Fault
    If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters flow fault. During flow fault, the pump stops, displays FLOW FAULT, pauses timing functions, and displays elapsed time or pump time. Auto-restart is attempted up to 5 times.
    Low Battery Fault
    Pumps stops, displays LO BATT, pauses timing functions, and displays elapsed time or pump time.
    Battery Test
    LCD indicator shows battery condition prior to sampling
    Time Display
    LCD displays up to 9999 minutes (6.8 days) at which point the display rolls over to 0. Displays include sampler run time in minutes for sampling period elapsed time, pump run time, or total elapsed time including delayed start
    Delay On
    Allows user to select minutes to delay start of test up to 9999 minutes (6.8 days)
    Timed Shutdown
    Allows user to select minutes of operation before automatic shutdown. Timed shutdown maximum is 9999 minutes (6.8 days)
    Timing Accuracy
    ± 0.05% (± 45 sec/day)
    Intermittent Sampling
    Programmable to allow user to extend short-term samples over an extended period of time to meet time-weighted average (TWA) requirements with a reduced number of samples. Elapsed time maximum is 9999 minutes (6.8 days), at which time the sample pump shuts down.
    Sampling Pause
    (Hold)
    Allows user to temporarily halt sampling without loss of timing data. Restart does not require resetting time.
    Multiple-tube Sampling
    Built-in constant pressure regulator allows user to take up to four simultaneous samples with rates up to 500 ml/min each (total combined flow 1350 ml/min maximum) using the optional adjustable tube holder accessory
    RFI/EMI Shielding
    Complies with requirements of EN 55022, FCC Part 15 Class B, EN 50082-1
    Frequency range of the radiated susceptibility test was 27 to 1000 MHz
    CE marked
    Additional Specifications
    Built-in particulate trap
    External see-through filter housing
    Tamper-resistant control cover
    Tamper-resistant compensation adjustments
    Built-in accessory mount
    Stainless steel belt clip
    External exhaust port
    All non-plastic parts made of corrosion-resistant materials
    RFI and EMI-shielded case
    Additional Features
    One-year warranty
    Ultra-quiet operation
    Pulsation-free flow
    Impact-resistant case